Installation and training is generally specifically designed so that the operators and users can learn how to use the equipment and quickly learn  through a series of procedures to produce acceptable results. However, our experience after several years of working with operators and researchers during installation and training, tells us that users learn more about the equipment and the technology of extrusion spheronization when things don’t immediately produce acceptable results. In this way they can learn what to do and how to react when things don’t go exactly as planned.

The installation of a Caleva bench-top Extruder 20 and Spheronizer MBS 250 at Diffucap Chemobras - Química e Farmacêutica followed this philosophy. An initial set of simple trials (as proposed by Caleva) were made during training. The diffucap scientists completed the trials using a Caleva bench top screen Extruder 20 and benchtop Spheronizer MBS 250. A summary of a small trial report  produced by the end user showing his view of the trials and the results obtained are shown below.

Initial Trial 1 with a mix of Avicel and water


Trials were completed using a mix of Avicel (PH101) and water. The samples were mixed in a small laboratory planetary mixer with water as a liquid binder as shown in the table below. Three tests were performed to check the functionality and to give a guide to the optimum amount of binder, to obtain acceptable pellets.

The parameters of the granulations used in all trials were not recorded. The mixture was extruded using the Extruder 20 with the MBS 250 Spheronizer with a speed range between 600 to 800 rpm. Samples were taken after 2 and 4 minutes of spheronization. The formulation details and results are shown below.

Products tested and results from trial 1

diffucap-trial-1-results

 

diffucap-trial-1-pellet-results-photograph
Figure 1 - Results showing pellets produced from an initial, intermediate step using a simple mix of Avicel and water that was considered unsatisfactory as the pellets had a poor level of “roundness”

The results were graded as being considered “satisfactory” or “not satisfactory”. The trial was not intended to provide a detailed and definitive optimum result.

Results showed that the initial trials with a simple formulation mix of Avicel and water needed some modification in order to obtain a satisfactory result. The equipment worked extremely well in demonstrating that it could produce acceptable results simply and easily, but that formulations did need some investigation and modification. This seems like a simple enough step to understand, but initially many operators that are new to the technique of extrusion spheronization need some experience to be able to get a “feel” for this basic idea.

 


Trial 2 with material “Premix B”

As with the previous trial series, a set of trials were completed using a different formulation to give a guide to the optimum binder requirement for a slightly more complex formulation suggested by the Diffucap researchers.

The liquid binder used was water, ethyl alcohol, or a mixture of the two. Mixing and extrusion parameters were not reported and the spheronizer was run between 600 and 800 rpm with sampling after 30 seconds and 60 seconds. In an initial test made with an inadequate amount of liquid the material in the spheronizer would not bind together and returned to a powder state very quickly. This test is not reported.

For the second test, a mixture of a substance “B” was mixed with povidone. This premix was granulated with a Povidone solution. Below are the tests performed, together with the grading for the results. Once again, using a slightly more complex formulation and by working through a series of "failures",  A satisfactory result was obtained quickly.

 

Products tested and results from trial 2

diffucap-trial-2-results

diffucap-trial-2-pellet-results-photograph
Figure 2 - Results showing pellets produced from trial number three using a simple mix of “Premix b” and water and/or alcohol as shown


Trial 3 with material C


Three tests were carried out to estimate the optimum amount of liquid binder with a different formulation. The premix (Material C) was granulated with a PVP solution. Mixing parameters and details of the formulation pre-mix were not given. Spheronizer parameter: 1000 rpm with sampling after 1 and 2 minutes. A summary procedure and of the results is shown below.

diffucap-trial-3-results
diffucap-trial-3-pellet-results-photograph
Figure 3 - Results showing pellets produced in the trial set 1 using a simple mix of Avicel and water showing an improvement in “roundness”



Trial 4 with material “Premix D”

Three tests with a more complex premix were completed to give a guide to the optimum amount of liquid binder required. Premix D (composition not disclosed) was granulated with the binder solution. Spheronizer parameter: 800rpm with sampling after 1 and 2 minutes.
Once again, the same procedure was used with initial trials not being considered as giving an acceptable result. A few additional trials with the formulation modified achieved an improved and an acceptable product was obtained.

diffucap-trial-4-results

diffucap-trial-4-pellet-results-photograph


Conclusion


The trials were not designed to give optimum results from the beginning. They were designed to show that operators could learn how to use the equipment quickly and work through series of procedures that would come closer and closer to delivering acceptable results. On the basis that generally during training operators learn more about equipment and technology of extrusion spheronization when things don’t produce acceptable results and when they do. The trials were good demonstration of this, and the customers being trained were happy both with the results obtained and the fact that they had an initial experience of solving formulation issues.

 

Remember:-

We can often learn more when things "fail" then when everything is perfect first time.


Relatório de Instalação e Utilização Extrusor e Esferonizador.pdf

 

Dr Steve Robinson
By Dr Steve Robinson
27 Mar 2020 in Caleva Installations and Training

Leave A Comment

 

 

 

Talk To Caleva